O2-A2 PERFORMANCE |
Output |
μA @ 20.9% O2 |
80 to 120 |
Response time |
t90 (s) from 20.9% to 0% O2 |
< 15 |
Zero current |
μA in N2 |
< 2.5 |
Linearity |
% O2 deviation @ 10% O2 |
0.6 |
|
|
|
O2-A2 Oxygen Sensor LIFETIME |
Output drift |
% change in output @ 3 months |
< 1 |
Operating life |
months until 85% original output of 20.9% O2 |
> 24 |
|
|
|
O2-A2 Oxygen Sensor ENVIRONMENTAL |
Humidity Sensitivity |
% O2 change: 0% to 95% rh @ 40℃ |
< 0.7 |
CO sensitivity |
% (change O2 reading) / % CO2 @ 5% CO2 |
0.1 |
Pressure sensitivity |
(% change of output)/(% change of pressure) @ 20kPa |
< 0.1 |
|
|
|
O2-A2 Oxygen Sensor KEY SPECIFICATIONS |
Temperature range |
℃ |
-30 to 55 |
Pressure range |
kPa |
80 to 120 |
Humidity range |
% rh non-condensing (0 to 99% rh short term) |
5 to 95 |
Storage period |
months @ 3 to 20℃ (store in sealed pot, open circuit) |
6 |
Load resistor |
Ω (recommended) |
47 to 100 |
Diameter |
mm (including label) |
20 |
Height |
mm (including foam ring) |
17.4 |
Weight |
g |
< 16 |